About Room 29
Room 29 is a semi-anechoic 3-meter chamber configured primarily for regulatory compliance testing. It is commonly used for FCC, ISED-Canada, CE Mark, and Global Regulatory services for unintentional radiators and digital devices. It is also used for measurements on intentional transmitters and receivers for wireless certifications. Tests results from this 3-meter chamber are accepted by regulators in North America and Europe, as well as by many globally agencies through our Mutual Recognition Agreements (MRAs). For automotive electronic components, this chamber is configured for EMC testing in accordance with CISPR 25 radiated emissions, ISO 11452-2 radiated immunity, ISO 11452-5 Strip-line immunity.
Room 29
Facility Description
FCC & ISED Canada Regulatory Emissions Testing
CE & UK European Regulatory Compliance
CISPR 25, ISO 11452-2 & 5 Automotive Testing
- Semi-Anechoic 3-Meter Chamber
- 32’x20’x18′ overall size
- 28’x16′ usable space
- 2meter diameter turntable
- 3300lbs (1500kg) turntable capacity
- Shielded adjacent monitoring room
- Configurable with a metallic bench ground plane.
Filtered Power:
AC: 115VAC, 208VAC, 480VAC 3-Phase, 30A
DC: 60A circuits
Other power arrangements are available upon request.
Cabling Interfaces:
25′ run from turntable into monitor room, below floor.
Fiber optics transceivers for Ethernet, CAN, I/O.